Home chevron_right Sensors chevron_right Gas Sensors chevron_right Toxic Gas Sensors 
Electrochemical Ethylene Oxide Sensor ME3-ETO
Electrochemical Ethylene Oxide Sensor

Electrochemical Ethylene Oxide Sensor - ME3-ETO

Low Power Consumption
High Precision
High Sensitivity
Wide Linear Range
Strong Anti-Interference Ability
Excellent Repeatability and Stability

Product Model: ME3-ETO

Product Introduction

Electrochemical Ethylene Oxide Sensor ME3-ETO Product Description

The ME3-ETO electrochemical gas sensor is based on the electrochemical principle, using the measured gas to undergo corresponding oxidation-reduction reactions on the sensor's working electrode and counter electrode, releasing charge, and forming a current through the external circuit. The current magnitude is proportional to the gas concentration, thereby achieving quantitative measurement of the target gas.


Electrochemical Ethylene Oxide Sensor ME3-ETOFeatures

Low Power Consumption

High Precision

High Sensitivity

Wide Linear Range

Strong Anti-Interference Ability

Excellent Repeatability and Stability


Electrochemical Ethylene Oxide Sensor ME3-ETOMain Applications

  • Widely suitable for ETO detection in industry and environmental protection

  • Technical Specifications

  • Product Model

    ME3-ETO

    Product Type

    Electrochemical Gas Sensor Element

    Product Package

    Plastic Package (ME3)

    Detection Range

    0—20ppm

    Measurement Limit

    100ppm

    Expected Life

    2 years

    Sensitivity

    1.8±0.3uA/ppm

    Resolution

    0.1ppm

    Operating Temperature Range

    ﹣20℃~+50℃

    Operating Pressure Range

    Standard atmospheric pressure ±10%

    Response Time (T90)

    ≤120S

    Humidity Range

    15% - 90% RH non-condensing

    Zero Drift (-20℃~+40℃)

    ≤4ppm

    Stability (/month)

    <2﹪

    Load Resistance (Recommended)

    10Ω

    Repeatability

    <2% output value

    Output Linearity

    Linear

    Bias Voltage

    +300mV