German Knick SE 620 Electrode Sensor SE 620
German Knick SE 620 Electrode Sensor

German Knick SE 620 Electrode Sensor - SE 620

Conductivity sensor for pharmaceutical design, featuring coaxial electrodes and an integrated temperature detector. Low surface roughness (<0.8 µm). Materials are physiologically harmless and meet FDA requirements. Steam sterilizable. Measurement results can be easily and reliably checked using the PortaSim simulator according to USP <645> standards.

Product Model: SE 620

Product Introduction

1. Description of SE 620 Electrode Sensor:

SE 620 Electrode sensorIt is a hygienic design 2-electrode sensor that meets pharmaceutical standards.

Conductivity sensor for pharmaceutical design, featuring coaxial electrodes and an integrated temperature detector. Low surface roughness (<0.8 µm). Materials are physiologically harmless and meet FDA requirements. Steam sterilizable. Measurement results can be easily and reliably checked using the PortaSim simulator according to USP standards.


2. Applications of SE 620 Electrode Sensor
Pure water and ultrapure water, water for injection (WFI), food industry, ion exchangers, reverse osmosis units, chip manufacturing


3. Technical Parameters of SE 620 Electrode Sensor

 Cell constant  0.01/cm
 Measuring range  0.001 ... 50 µS/cm
 Material  Cell and electrodes: stainless steel 1.4435, electropolished;
Insulator and O-rings (plastics), FDA-listed
 Roughness  < 0.8 µm
 Temperature detector  Pt 1000
 Temperature  0 ... 135 °C (steam-sterilizable)
 Pressure  16 bar at 25 °C, 9 bar at 60 °C
 Process connection  Clamp DN 25
 Sensor cap  VP (VarioPin)

 

4. Dimensions of SE 620 Electrode Sensor