Dust Particle Counter - 3886
- Supports 5-channel testing (0.3, 0.5, 1.0, 3.0 & 5.0 μm)
- Added GB MODEL test function, test reports comply with GB/T16292-2010 standard
- Built-in delay test time, user-adjustable to avoid human interference
- Built-in flow sensor ensures high-precision particle counting
- High-efficiency data storage and recording capability (stores 500 data points)
- Compact structure, easy to carry
- Can be used as a multi-point monitoring system with cascade connection
- Can be connected to an external computer for remote monitoring
Product Introduction
Dust Particle Counter 3886 Product Overview:
- Supports 5-channel testing (0.3, 0.5, 1.0, 3.0 & 5.0 μm)
- Added GB MODEL test function, test reports comply with GB/T16292-2010 standard
- Built-in delay test time, user-adjustable to avoid human interference
- Built-in flow sensor ensures high-precision particle counting
- High-efficiency data storage and recording capability (stores 500 data points)
- Compact structure, easy to carry
- Can be used as a multi-point monitoring system with cascade connection
- Can be connected to an external computer for remote monitoring
Dust Particle Counter 3886 Product Features:
| Technical Parameters |
|
| Product Name |
Dust Particle Counter 3886 |
| Particle Size Range |
0.3 、0.5、 1.0、 3.0、 5.0μm |
| Light Source |
Laser Diode |
| Counting Efficiency |
Refer to JIS B9921 |
| Zero Count |
Refer to JIS B9921 |
| Coincidence Loss |
Less than 5% at 2,000,000 particles per cubic foot |
| Flow Rate |
0.1cfm(2.83 l/min) |
| Sampling Time |
1sec~99min59sec |
| Sampling Frequency |
1 to 99 times or continuous |
| Count Alarm |
In the range of 1 to 70,000,000 counts |
| Test Mode |
Single/Repeat/Continuous/Calculation/Remote |
| Display |
20 characters, 4-line LCD |
| Error Indication |
Over count, laser power attenuation, over calibration flow (±10%), low battery |
| Interface |
RS-232C or RS-485 (selectable in menu), RJ-11 to N.B. RS-485 for cascade connection |
| Communication Speed |
9600bps |
| Buffer |
500 data points (in calculation mode, one test counts as 4 data points) |
| Power Supply |
4 AA batteries, Ni-MH battery (4.8V-1.6Ah) or AC adapter (input 100-240V). The battery cannot be charged via AC adapter; use the included dedicated charger. |
| Battery Operating Time |
Approximately 3 hours of continuous testing (Ni-MH battery) |
| Dimensions |
115(W)×70(D)×211(H)mm |
| Weight |
Approximately 980g (excluding batteries) |
| Operating Environment |
Ambient temperature: 10-35°C |
| Standard Accessories |
AC adapter, filter, tubing, user manual, charger, rechargeable battery |
| Optional Accessories |
Printer, printer cable, temperature/humidity probe, wind speed probe, extension rod for wind speed probe, portable protective case, tripod |
Comparison Table of Main Parameters for Dust Particle Counter Models
| Model & Main Parameters |
3887L |
3887 |
3887C |
3887D |
3886 |
3900 |
| Particle Size Range |
0.5 、 5.0um |
0.3 、 0.5 、 5.0um |
0.3 、 0.5 、 1.0 、 3.0 、 5.0um |
0.3 、 0.5 、 1.0 、 3.0 、 5.0 、 10.0um |
||
| Operating Interface |
Chinese |
English |
Chinese |
Chinese |
English |
English |
| Flow Rate |
0.1cfm ( 2.83L /min ) |
1.0cfm (28.3L/min) |
||||
| Sampling Time |
1sec ~ 99min59sec |
1sec ~ 24h |
||||
| Data Storage |
3000 |
10000 |
7000 |
7000 |
500 |
5000 (CF card) |
| Test Mode |
Single, GB |
Single, Repeat, Continuous, Calculation, Remote, GB, ISO (3887C, D only) |
General, Statistical, Standard (various test standards) |
|||