Product Introduction
1. Description of UK Detectronic Water Quality Sensor MMs-d010:
The ORAKEL series residual chlorine analyzers, residual chlorine controllers, and residual chlorine monitors are membrane devices that are insensitive to pH changes, require no reagents, are extremely stable, and reduce maintenance and lifecycle costs.
The thin-film amperometric chlorine sensor incorporates a third reference electrode to eliminate zero drift. Its unique design means pH correction is typically not needed at all, completely eliminating reagents.
In addition to advanced potentiostatic amperometric free chlorine and total chlorine sensors, the ORAKEL series residual chlorine analyzers have all the features you need, and more.
2. Applications of UK Detectronic Water Quality SensorMMs-d010
The ORAKEL chlorine sensor range is particularly suitable for applications where reliability and ease of use are of utmost importance.
Selection of ORAKEL Chlorine Sensors
Free Chlorine
Total Chlorine
Free Chlorine 'Zero'
Water Treatment Applications
Chlorine Dosing
Paper Mills
Cooling Towers
Remote Sites
Food Preparation
Chlorination
ORAKEL chlorine sensors can be installed in various auxiliary flow cells and self-cleaning units.
3.UK Detectronic Water Quality SensorMMs-d010 Parameters:
| Type |
Membrane covered potentiostatic chronoamperometric three-electrode system |
| Measurement |
Total chlorine or free chlorine |
| Range |
0.01-2mg/l, 0.01-5mg/l, 0.01-10mg/l, 0.01-20mg/l and 0.5-200mg/l (free only). |
| Resolution |
0.01mg/l (ppm) (0.1 on 0-200mg/l ranges). |
| Reproducibility |
Better than ±0.05mg/l. |
| Stability |
-1% per month (without calibration). Working electrode: gold. Counter electrode: stainless steel. Reference electrode: silver/silver halide |
| Membrane Material |
Membrane Material. |
| Flow Rate |
Approximately 0.5l/minute (minimum 0.25l/minute). |
| Temperature Range |
0 to <45°C. |
| Temperature Compensation |
Automatically by an integrated thermistor. |
| pH Range |
pH 4 up to pH 9. |
| First Polarisation Time |
2 hours. |
| Re-Polarisation Time |
30 minutes. |
| Response Time |
T90: approximately 120 seconds. |
| Zero-Point Adjustment |
Not necessary. |
| Calibration |
Manual using electrochemical test kit or DPD test kit. |
| Housing Material |
PVC, silicone and polycarbonate stainless steel. |
| Dimensions |
Diameter: approximately 25mm. Length: 175mm. |
| Maintenance Intervals |
Membrane: once a year (depending on water quality). Electrolyte: once a year (depending on water quality). |
| Interferences |
High levels of other oxidants such as Ozone and Chlorine Dioxide |