Alphamoisture Online Intelligent Dew Point Meter - DS-3000
⊙ User on-site multi-point calibration function.
⊙ Multi-unit display: can simultaneously display ℃ (DP), PPMv.
⊙ Wide range: selectable measurement range of -110 to +20℃ DP.
⊙ High accuracy: +/-1℃ measurement accuracy.
Product Model: DS-3000
Product Introduction
I. Description of Alphamoisture Online Intelligent Dew Point Meter DS-3000:
The DS-3000 dew point meter is a single-channel online precision dew point analyzer, suitable for monitoring various industrial gases and clean compressed air. The DS-3000 dew point analyzer combines advanced high-capacitance aluminum oxide sensors with advanced digital circuit technology for precise dew point detection.
The combination of advanced display technology and comprehensive dew point sensor technology makes the dew point analyzer highly flexible and adjustable. The menu-guided operation mode facilitates user adjustments and operations. The display unit of the analyzer can be flexibly set at the factory, with options including ℃, ℉, ppm(v), g/m3, or lb/MMSCF.
II. Features of Alphamoisture Online Intelligent Dew Point Meter DS-3000:
User on-site multi-point calibration function.
Multi-unit display: can simultaneously display ℃ (DP), PPMv.
Wide range: selectable measurement range of -110 to +20℃ DP.
High accuracy: +/-1℃ measurement accuracy.
Dual alarms: two adjustable alarm outputs.
The distance between the sensor and the secondary meter can be up to 1 kilometer. Secondary meter
III. Parameters of Alphamoisture Online Intelligent Dew Point Meter DS-3000:
Model: DS-3000.
Display: large screen LCD display, alarm status display.Display units: ℃ (DP), PPMv.
Resolution: 0.1℃ (or 0.01ppm).
Power supply: 220V AC @ 50Hz.
Input signal: dedicated SHAW sensor
Output signal: 4~20mA current, 0-5V voltage, USB serial output.
Alarm output: 2 adjustable relay alarms, 2A@250V AC.
Dimensions: 95 (W) x 48 (H) x 110 (D) mm.
Weight: 0.5Kg.
Mounting: panel mounting, cutout size: 92 x 45 mm.
Protection rating: IP65.
Dew Point Sensor
Model: Sensor.
Principle: aluminum oxide capacitance method.
Pressure: 0~50Bar (standard), optional 200Bar.
Explosion-proof: intrinsically safe ATEX certified.
Cable: NPN connector, maximum distance 1000 meters, standard 2 meters.
System Parameters
Range: standard -80~0℃ DP or 0~100PPMv, other ranges optional.
Accuracy: ±2℃ (DP).
Signal isolation: power supply and internal circuits are fully isolated; input and output are also fully isolated.
Gas connection: sensor comes with a sensor cover, using 1/4" stainless steel compression fittings.
Explosion-proof: can be used in explosive atmospheres with an isolation barrier, explosion-proof rating: EEx ia IIC T6.
Calibration: on-site user automatic calibration, multi-point calibration by distributors.
Operating temperature: -10~+50℃, storage temperature: -50~+70℃.
Operating humidity: 10~90%RH (non-condensing).
Recommended Products