UK Alphasense Micro Chlorine Gas Sensor - CL2-D4
Performance: Sensitivity 10ppmCL2, nA/ppm -200~-450
Response time t90 from zero to 10ppmCL2 (s) <35
Zero current: ppm in zero air < ±0.8
Resolution: average noise (ppm) < 0.1
Range: CL2 warranty detection range (ppm) 20
Product Introduction
I. Main Features of Micro Chlorine Gas Sensor CL2-D4
Sensitivity: -200~-450nA/ppm
Response time: <35s
Linearity range: +/-0.5ppm
Overload: 60ppm
Resolution: 0.1ppm
Dimensions: Φ14.5*8.3
Service life: 2 years
Storage period: 6 months
Operating temperature: -20~50℃
Operating humidity: 15~90%RH
Load resistor: 33Ω
II. Typical Applications of Micro Chlorine Gas Sensor CL2-D4
Wastewater treatment, toxic gas measurement in industrial environments
III. Diagram of Micro Chlorine Gas Sensor CL2-D4
Micro Chlorine Gas Sensor
| Performance |
Sensitivity |
10ppmCL2,nA/ppm |
-200~-450 |
|
| |
Response time |
t90 from zero to 10ppmCL2 (s) |
<35 |
|
| |
Zero current |
ppm in zero air |
|
< ±0.8 |
| |
Resolution |
Average noise (ppm) |
< 0.1 |
|
| |
Range |
CL2 warranty detection range (ppm) |
|
20 |
| |
Linearity |
Full scale ppm error, linear at 0-5ppm |
±0.5 |
|
| |
Overload |
Maximum ppm for stable response to gas pulses |
60 |
|
| |
|
|
|
|
| Lifetime |
Zero drift |
ppm change per year in laboratory air |
|
nd |
| |
Sensitivity drift |
Percentage change per year in laboratory air |
nd |
|
| |
Operating life |
Time until output drops to 80% of original signal (months) (warranty 24 months) |
> 24 |
|
| |
|
|
||
| Environment |
Sensitivity at -20°C |
At 10ppmCL2, (output at -20°C/output at 20°C)% 80-110 |
||
| |
Sensitivity at 50°C |
At 10ppmCL2, (output at 50°C/output at 20°C)% |
95-125 |
|
| |
Zero at -20°C |
Reference ppm change from 20°C |
|
-0.4~0.4 |
| |
Zero at 50°C |
Reference ppm change from 20°C |
|
0~0.5 |
| |
|
|
|
|
| Cross |
H2S |
Percentage sensitivity to 20ppmH2S |
< -20 |
|
| Sensitivity NO2 |
Percentage sensitivity to 10ppmNO2 |
<80 |
||
| |
NO |
Percentage sensitivity to 50ppmNO |
<0.5 |
|
| |
SO2 |
Percentage sensitivity to 20ppmSO2 |
<1 |
|
| |
CO |
Percentage sensitivity to 400ppmCO |
<0.1 |
|
| |
H2 |
Percentage sensitivity to 400ppmH2 |
<0.1 |
|
| |
C2H4 |
Percentage sensitivity to 400ppmC2H4 |
<0.1 |
|
| |
NH3 |
Percentage sensitivity to 20ppm NH3 |
<0.1 |
|
| |
|
|
|
|
| Key |
Temperature range |
℃ |
|
-20 ~ 50 |
| Parameters |
Pressure range |
Kpa |
|
80-120 |
| |
Humidity range |
%rh |
|
15-90 |
| |
Storage life |
Sealed storage life at 3-20℃ (months) |
|
6 |
| |
Load resistor |
Ω (recommended) |
|
33 |
| |
Weight |
grams |
|
<2 |