UK Alphasense Micro Chlorine Gas Sensor CL2-D4
UK Alphasense Micro Chlorine Gas Sensor
UK Alphasense Micro Chlorine Gas Sensor

UK Alphasense Micro Chlorine Gas Sensor - CL2-D4

Performance: Sensitivity 10ppmCL2, nA/ppm -200~-450
Response time t90 from zero to 10ppmCL2 (s) <35
Zero current: ppm in zero air < ±0.8
Resolution: average noise (ppm) < 0.1
Range: CL2 warranty detection range (ppm) 20

Product Model: CL2-D4

Product Introduction

 

I. Main Features of Micro Chlorine Gas Sensor CL2-D4
Sensitivity: -200~-450nA/ppm
Response time: <35s
Linearity range: +/-0.5ppm
Overload: 60ppm
Resolution: 0.1ppm
Dimensions: Φ14.5*8.3
Service life: 2 years
Storage period: 6 months
Operating temperature: -20~50℃
Operating humidity: 15~90%RH
Load resistor: 33Ω

II. Typical Applications of Micro Chlorine Gas Sensor CL2-D4
Wastewater treatment, toxic gas measurement in industrial environments


III. Diagram of Micro Chlorine Gas Sensor CL2-D4


Micro Chlorine Gas Sensor

Performance

Sensitivity

10ppmCL2,nA/ppm

-200~-450


Response time

t90 from zero to 10ppmCL2 (s)

<35


Zero current

ppm in zero air


< ±0.8


Resolution

Average noise (ppm)

< 0.1


Range

CL2 warranty detection range (ppm)


20


Linearity

Full scale ppm error, linear at 0-5ppm

±0.5


Overload

Maximum ppm for stable response to gas pulses

60






Lifetime

Zero drift

ppm change per year in laboratory air


nd


Sensitivity drift

Percentage change per year in laboratory air

nd


Operating life

Time until output drops to 80% of original signal (months) (warranty 24 months)

> 24




Environment

Sensitivity at -20°C

At 10ppmCL2, (output at -20°C/output at 20°C)% 80-110


Sensitivity at 50°C

At 10ppmCL2, (output at 50°C/output at 20°C)%

95-125


Zero at -20°C

Reference ppm change from 20°C


-0.4~0.4


Zero at 50°C

Reference ppm change from 20°C


0~0.5





Cross

H2S

Percentage sensitivity to 20ppmH2S

< -20

Sensitivity NO2

Percentage sensitivity to 10ppmNO2

<80


NO

Percentage sensitivity to 50ppmNO

<0.5


SO2

Percentage sensitivity to 20ppmSO2

<1


CO

Percentage sensitivity to 400ppmCO

<0.1


H2

Percentage sensitivity to 400ppmH2

<0.1


C2H4

Percentage sensitivity to 400ppmC2H4

<0.1


NH3

Percentage sensitivity to 20ppm NH3

<0.1






Key

Temperature range

℃


-20 ~ 50

Parameters

Pressure range

Kpa


80-120


Humidity range

%rh


15-90


Storage life

Sealed storage life at 3-20℃ (months)


6


Load resistor

Ω (recommended)


33


Weight

grams


<2