Product Introduction
Rotronic Temperature and Humidity Transmitter XB32 Description:
The XB OEM transmitter includes a wired probe and PCB circuit board. Due to its compact size, high accuracy, and multiple options, the XB transmitter can meet any user needs.
Rotronic Temperature and Humidity Transmitter XB32Features:
- Probe application range: depends on probe -100 to 200 °C
- PCB temperature range: -40...85 °C
- Compatible with multiple types of probes
- Adjustable analog output signal range
- Simulator mode
- Direct 4-wire Pt100 connection (optional)
- Power Supply
- Low voltage supply 3/4-wire (XB3x)
- Low voltage supply 2-wire (XB20)
- Signal Output
- Current output
- Voltage output
- Version
- PCB with cable probe
- Probe
- PPS and stainless steel probes
- Probe diameter: 15 mm or 25/15 mm
- Probe length up to 700 mm
- Cable length 2 to 5 meters
- Output Parameters
- Humidity & Temperature
- Dew point/frost point and temperature & humidity
Rotronic Temperature and Humidity Transmitter XB32Applications:
OEM transmitters in environmental test chambers, incubators, monitoring of industrial production processes.
Rotronic Temperature and Humidity Transmitter XB32 Parameters:
| Accuracy at 23 ±5 °C |
±1.0 %rh / ±0.2°C |
| Operating Range |
-40~85 °C / 0~100 %rh |
| Integrated Probe Measurement Range |
-100~200 °C / 0~100 %rh |
| Long-term Stability |
<1 %rh / year |
| Sensors |
Hygromer IN-1 |
| Temperature Sensor |
PT100 1/3 DIN Class B |
| Maximum Air Speed |
40 m/s, with filter |
| Calculated Values |
Dew point or frost point |
| Number of Probe Inputs |
1 |
| Software Updates |
Via internet and HW4 software |
| Circuit Type |
2-wire or 3/4-wire |
| Maximum Cable Length |
Passive cable up to 5m, active cable up to 100m |
| Probe Material |
Depends on probe |
| Service Interface |
Mini USB (UART) |
| FDA / GAMP Compatibility |
Compatible with FDA 21 CFR Part 11 and GAMP |