Home chevron_right Sensors chevron_right Dissolved Oxygen Sensors 
German Knick Optical Oxygen Sensor SE 740
German Knick Optical Oxygen Sensor

German Knick Optical Oxygen Sensor - SE 740

Features
● Flow-independent measurement principle
● No polarization required
● Replaceable sensor element
● Integrated wear monitoring
● Stainless steel 1.4435, inspection certificate 3.1, Ra < 0.4 µm
● Gaskets compliant with FDA and USP VI requirements
● Temperature range: –10 … 130 °C, measurement: 0 … 80 °C
● Pressure range: 12 bar
● Measurement range: 0.05 ... 300 % saturation
● Reliable digital communication via Stratos Evo

Product Model: SE 740

Product Introduction

1. Description of SE 740 Optical Oxygen Sensor

Optical oxygen measurement based on fluorescence quenching technology

Flow-independent, fast-responseoptical oxygen sensor。
Ideal for measurements in demanding hygienic environments.
Steam sterilizable, autoclavable, and CIP supported.


2. Features of SE 740 Optical Oxygen Sensor
● Flow-independent measurement principle
● No polarization required
● Replaceable sensor element
● Integrated wear monitoring
● Stainless steel 1.4435, inspection certificate 3.1, Ra < 0.4 µm
● Gaskets compliant with FDA and USP VI requirements
● Temperature range: –10 … 130 °C, measurement: 0 … 80 °C
● Pressure range: 12 bar
● Measurement range: 0.05 ... 300 % saturation
● Reliable digital communication via Stratos Evo


3. Technical Parameters of SE 740 Optical Oxygen Sensor

 Measuring range  4 ppb / 0.05 ... 300 % saturation
 Resolution  0.1 ppb
 Response time at 25 °C (air –> N 2 )  98 % full scale < 30 s
 Pressure range P rel  –1 …12 bar
 Temperature range  –10 … 130 °C
measurements: 0 … 80 °C
 O-ring material  silicone and FKM (Viton)
 Process-wetted sensor parts  stainless steel 1.4435
 Temperature detector  NTC 22 kohms
 Process adaptation  Pg 13.5
 Sensor cap  M12 (8 pins)


4. Dimensions of SE 740 Optical Oxygen Sensor