Home chevron_right Sensors chevron_right Dissolved Oxygen Sensors 
German Knick SE 707 Dissolved Oxygen Sensor SE 707
German Knick SE 707 Dissolved Oxygen Sensor

German Knick SE 707 Dissolved Oxygen Sensor - SE 707

High-resolution measurement of low oxygen concentrations, simple, efficient, and low-cost sensor calibration in ambient air. No calibration gas required. High signal stability even in environments with high CO2 partial pressure. Materials comply with FDA requirements and have an extremely smooth polished surface. Modular design simplifies maintenance. Sterilizable, autoclavable, CIP-compatible. Also suitable for hazardous areas.

Model: SE 707

Product Introduction

1. Description of SE 707 Dissolved Oxygen Sensor:

SE 707 Dissolved Oxygen SensorsReliably measures trace oxygen even under sterilization and hygienic conditions; stable signal, robust and durable.

High-resolution measurement of low oxygen concentrations, simple, efficient, and low-cost sensor calibration in ambient air. No calibration gas required. High signal stability even in environments with high CO2 partial pressure. Materials comply with FDA requirements and have an extremely smooth polished surface. Modular design simplifies maintenance. Sterilizable, autoclavable, CIP-compatible. Also suitable for hazardous areas.

 

2. Applications of SE 707 Dissolved Oxygen Sensor
Beverage filling (e.g., milk, beer). Measurement in boiler feed water (detection limit 1 ppb)


3. Technical Parameters of SE 707 Dissolved Oxygen Sensor

 Range   pO 2 < 1200 mbar
 Detection limit  1 ppb (3 ppb in liquids containing CO 2 )
 Accuracy  1 % + 1 ppb (1 % + 3 ppb in liquids containing CO 2 )
 Response time at 25 °C (air –> N 2 )  98 % full scale < 90 s
 Signal in ambient air  290 … 500 nA
 Residual signal  in O 2 -free medium: ≤ 0.01 % of signal in ambient air
in CO 2 : ≤ 0.03 % of signal in ambient air
 Flow dependence  ≤ 5 %
 Perm. pressure range (measurement)  0.2 … 6 bar absolute
 Mechanical pressure resistance  max. 12 bar absolute
 Perm. temp range (measurement)  0 … 80 °C
 Temp range (stability)  –5 … +121 °C (steam sterilizable/autoclavable)
 O-ring material  silicone (FDA&USP Class VI tested)
 Membrane material  PTFE/silicone/PTFE, FDA comp. (reinforced w. steel mesh)
 Process-wetted sensor parts  1.4404 stainless steel (material certificate 3.1B)
 Surface roughness  N 5 (RA < 0.4 µm)
 Temperature detector  NTC 22 kohms
 Interior body incl. sensing electrodes  replaceable (ZU 0568)
 Sensor cap  VP (VarioPin)
 ATEX Certificate  II 1/2 GD Ex ia IIC T6 / T5 / T4 / T3
 Certificates  3A, 3.1B (EN 10204/3.1.B)

 

4. Dimensions of SE 707 Dissolved Oxygen Sensor